  1. Inspection, cleaning and disinfection.
  2. Tightness test and change of O'ring if necessary.
  3. Calibration in 3 volumes and 10 dispensings per volume.
  4. Counter adjustment if necessary.
  • DSCP16CSB
  • Available upon request
  • Free
Qty:
Factory verification required

  Terms and conditions of sale

Be sure to read our terms and conditions of payment, shipping, returns and product warranty.

Pipette Calibration Service

On arrival of the pipette at our workshop, a complete initial inspection is carried out, followed by the cleaning and disinfection process, paying special attention to the most susceptible parts, such as the tip holder and piston. 

Once this has been done, the mechanism is checked to ensure that it is working correctly and the hermeticity test is carried out, where it is observed whether the pipette is vacuuming correctly. If it does not, the seals are replaced, and the test is repeated to ensure that the pipette is suitable in terms of leak-tightness.

A series of pre-weighing tests are carried out to check that the pipette is working correctly and it is then taken to the calibration room to be tempered.

When checking the volumes, if the pipette does not fall within the range described by the manufacturer or, failing that, the range of the ISO standard, the meter is adjusted and the corresponding certificate is issued. 

All of this is done in compliance with the gravimetric method, in temperature, pressure and humidity, as defined by the ISO-8655 Standard.

DSCP16CSB

Specific References